COMMERCIAL ITEM DESCRIPTION A-A-51019
Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the Government’s request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate.
Inspection. Inspection, as used herein, is defined as both examination (such as visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item.
Tests. Tests shall be conducted in accordance with commercial practice to determine compliance with specification requirements. Where feasible, the same sample shall be used for determination of two or more test characteristics.
Sampling for tests. Sampling for tests, including dimensional test, shall be conducted in accordance with MIL-STD-105 and as specified herein. The unit of product for test purpose shall be one Catheter to Luer Syringe Adapter.
Testing requirements. Adapters shall be subjected to the following tests.
Corrosion-resistance. Corrosion-resistance of adapters will be determined by the salt-spray and boil-test methods in Federal specification GG-526.
For the salt-spray test the exterior of the adapter will be determined by the significant surface. Each piece of the adapter must be tested separately.
Dimensional testing. Test adapters in accordance with Section 7 of ANSI/HIMA MD70.1. “Dimensional Test for Luer-Lock Fittings”.
Performance testing. Test adapters as follows. Prepare for testing by heat-flaring a size 6 French catheter with a 45o flaring mandrel. Check the catheter for proper flare depth and angle. Connect the adapter and close the catheter with a suitable pinch-clamp. Perform the test specified in section 8 of ANSI/HIMA MD70.1, “Performance Tests for Luer-Lock Fittings”.
Contractor certification. The contractor shall certify that the product offered meets the salient characteristics of this description and conforms to the producers’ own drawings, specifications, standards, and quality assurance practices. The Government reserves the right to require proof of such conformance prior to first delivery and thereafter as may be otherwise provided for under the provisions of the contract.