COMMERCIAL ITEM DESCRIPTION A-A-51408
Thermal stability. The colon tube, if supplied non-sterile, shall pass the thermal stability test listed below.
Thermal stability test. Sample colon tubes will be subjected to cyclic steam autoclaving at 250 ± 5O F, 1 psig. A total of 10 cycle, the colon tube will be examined for any sign of brittleness, blistering, swelling, peeling, or cracking.
Workmanship. The colon tube shall be free from any defects that detract from its appearance or impair its serviceability.
Unit. Package (PG). One package containing 50 individually packaged colon tubes, as specified, constitutes one unit.
Quality Assurance Provisions.
Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contact or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements.
Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the Government’s request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate.
Inspection. Inspection, as used herein, is defined as both examination (such as visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item.
Tests. Tests shall be conducted to determine compliance with specification requirements. Where feasible, the same sample shall be used for the determination of two or more test characteristics.
Sampling for tests. Sampling for tests shall be conducted in accordance with MIL-STD-105 and as specified herein. The unit of product for test purposes shall be one colon tube.