A-A-51545B
Sampling for examination. Sampling for examination shall be conducted in accordance with MIL-STD-105, with an AQL of 1.0 (percent defective) and an inspection level of II. The unit of product examination purposes shall be one humidifier as specified in this description.
Defects.
Humidifier not bubble type.
Absence of relief valve.
Head not easily removed/installed.
Reservoir marking not as specified.
Aerosol outlet does not accept 3/16 inch and 1/4 inch ID tubing.
Components not free of cracks, burrs, pits, sharp edges, or corrosion.
Components not free of pits or unintended mold marks.
Sampling for tests. Sampling for the following tests shall be conducted in accordance with MIL-STD-105 with an AQL of 1.5 (percent defective) and an inspection level of S-1:
Overpressure relief
Warning signal
(aerosol circuitry occlusion)
Flow rates
Fluid outputs
Sampling for the reservoir capacity test shall also be in accordance with MIL-STD-105, with no defects permitted.