A-A-51552A
Workmanship. Airways shall be free from warpage, weld-lines, shorts, pits, pimples, fish-eyes, blisters, shrink-marks, blooms and any other defects which detract from their appearance or impair their serviceability. Definitions of defects listed above shall be in accordance with ASTM D883, Standard Definitions of Terms Relating to Plastics.
Airways shall be supplied clean and dry.
Unit. Package (PG). One package containing thirty (30) size 10 individually packaged airways or fifty (50) size 6 individually packaged airways as specified, constitutes one unit.
Quality Assurance Provisions.
Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contact or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements.
Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the Government’s request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate.
Inspection. Inspection, as used herein, is defined as both examination (such as visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item.
Sampling for inspection. Sampling for inspection of all material properties shall be conducted in accordance with ASTM D1898, except that the Acceptable Quality Level (AQL) and Inspection Level shall be 1.0 percent and S-2 per MIL-STD-105.