A-A-51597A
15 June 1990
SUPERSEDING
A-A-51597
22 August 1986
COMMERCIAL ITEM DESCRIPTION
ELECTRODE, ELECTROCARDIOGRAPH
The General Services Administration has authorized the use of this commercial item description.
This commercial item description covers a disposable pregelled electrocardiograph electrode suitable for cardiac monitoring.
Salient characteristics:
The electrocardiograph electrode shall be a low impedance, disposable, pregelled, adhesive foam, electrode with silver/silver chloride sensor and a snap connector.
Construction shall be of soft, pliable materials to provide maximum patient comfort. Adhesive shall be non-irritating to the skin.
Electrode shall be approximately mm in diameter.
Electrode shall permit low DC offset and minimum noise artifact recording when applied properly.
Electrode shall comply with the applicable requirements of American National Standard for Pregelled ECG disposable Electrodes (ANSI/AAMI EC12).
Electrode shall have a minimum 24 month shelf life. Not less than 20 months of the expiration dating period shall remain at time of delivery to the Government.
Workmanship. The electrocardiograph electrode shall be free from defects which detract from their appearance or impair its serviceability.
DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited.
For Parts Inquires call Parts Hangar, Inc (727) 493-0744
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business