A-A-51602B
Each inner canister shall be provided with a snap seal lid which prevents leakage and spilling during aspiration and disposal. Collapsible items may have an integral lid sealed to the bag.
The canister assembly shall be provided with a shut-off valve to protect suction source and shall have the following ports:
* Vacuum port. The vacuum port shall have a built in filter to minimize bacterial spray into the vacuum system.
* Patient port. The patient port shall include a right angle swivel connector for elimination of patient tube kinking.
* Accessory port. The accessory port shall incorporate a pour spout designed to allow the canister contents to be emptied in an area isolated from patients and staff.
The suction canister assembly shall be capable of being exposed to an implosion vacuum of 690 mm Hg without degradation (implosion vacuum measured in occluded inner canister), and shall be capable of operating while being exposed to vacuums up to 500 mm Hg as measured at or near the patient end of 48 inch long patient tube with catheter end occluded and having 300 to 500 ml of water in the canister.
The outer canister shall be autoclavable up to a temperature of 250°F for at least one sterilization cycle.
Workmanship. The suction canister assembly with the 100 inner canisters shall be free from defects which detract from their appearance or impair their serviceability.
Unit. Package (PG). One package containing one hundred disposable inner canisters along with one reusable outer canister, as specified, constitutes one unit.
Quality Assurance Provisions.
Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contact or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed
For Parts Inquires call Parts Hangar, Inc (727) 493-0744