Controls.
Compression force. The resuscitator shall have an adjustable compression force range of 0 to 145 pounds (0 to 65.8 Kg) when operated from a supply pressure of 50 psi.
Compression stroke. The compression stroke shall be adjustable from 0 to 3/4 inches (o to 1.90 cm).
Ventilation pressure. The unit shall have an adjustable ventilation pressure range of 5.0 to 60.0 cm H2O.
Compression frequency. The compression frequency shall be adjustable from 40 to 120 cycles/minute.
Chest time pressure. The chest time pressure shall be adjustable from 30 percent to 70 percent of compression cycle.
Ventilation duration. The unit shall have an adjustable ventilation period from 0.5 to at least 2.0 seconds.
Components.
The components of the unit shall at least include the following:
a. Carrying/storage case with hook and pile straps for holding backboard.
b. Transport model backboard with straps.
c. Mobile oxygen carrier for “D” size cylinders.
d. DISS pipeline adapter.
Performance.
Ventilation output. The unit shall have a ventilation output of 140 L/minute, minimum at 20 cm H2O backpressure.
Materials.
Metal parts. Alloy metal parts shall be fabricated of a suitable metal alloy that has been properly hardened and passivated.
Plastic parts. All plastic parts shall be fabricated of materials which are in accordance with USP Class IV, Plastic.
Instructions. Each unit shall include one (1) each user’s manual and two (2) each repair manual.