A-A-53005
TABLE I. Sampling for examination.
|
Inspection level |
AQL (defects per hundred units) |
For visual examination |
|
|
Major defects |
II |
1.0 |
Total defects (Major and minor combined) |
II |
4.0 |
For dimensional examination |
S-1 |
2.5 |
Examination. The nebulizer bottle shall be examined to determine compliance with all requirements contained in this specification.
Classification of defects. Examination shall be conducted in accordance with the classification of defects listed in table II. Examination shall not be restricted to the classified possible defects below.
TABLE II. Categories and defects.
Categories |
Defects |
Major |
|
101 |
Any component part missing broken, or otherwise unusable. |
102 |
Oxygen dilution selector dial does not permit necessary selection range. |
103 |
Nebulizer adapter not non-ferrous. |
104 |
Bottle not capable of being temperature controlled in a suitable heating device. |
Minor |
|
201 |
Outside of bottle not clean. |
Contractor certification. The contractor shall certify that the product offered meets the salient characteristics of this description and conforms to the producers’ own drawings, specifications, standards, and quality assurance practices. The Government reserves the right to require proof of such conformance prior to first delivery and thereafter as may be otherwise provided for under the provisions of the contract.