A-A-54039
Litter mounting attachment. The self-contained apparatus shall include a litter mounting attachment consisting of four D-rings, two attached to the case front and two attached to the case rear, and two 24 inch long hook and pile/nylon webbing attachment straps. The straps shall be shipped in the carrying case.
Separate stepdown transformer. The unit shall include a separate stepdown transformer as a separate component not required to be in the carrying case, but shipped in the same shipping package. The transformer shall accept 220-250 volts, 50/60 Hz and provide one half output voltage for the suction apparatus. The transformer shall include a hospital grade, 110 volt receptacle to accept the suction apparatus’ specified male line cord. The transformer shall include 3 posts in a compartment with a wire strain-relief attachment for SJT-16/3 wire. The three posts shall be wired to accept any worldwide wiring scheme required to connect a cord for 220-250 volts, single phase AC service. The separate stepdown transformer shall be rated for continuous standby use. The transformer shall comply with UL-1012, including testing.
Power sources. The self-contained apparatus shall be operational from external 12-14 volts DC vehicle generating system rated at 120 amperes at 14 volts, external 120 amp-hour heavy-duty 12 volt battery, 21 volt internal battery, and 110-125 volt, 50/60/400 Hz power sources.
In addition, the separate transformer shall be employed as a power source for the apparatus by reducing 220-250 volts, 50/60 Hz to 110-125 volts.
Operating temperature range. The operating temperature range shall include -4°F to +115°F. The apparatus shall meet the requirements in accordance with ASTM F960 using the internal battery, the 120 amp hour external battery, and 110 volts AC 50, 60, and 400 Hz.
For Parts Inquires call Parts Hangar, Inc (727) 493-0744
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business