A-A-54039
Litter mounting attachment. The self-contained apparatus shall include a litter mounting attachment consisting of four D-rings, two attached to the case front and two attached to the case rear, and two 24 inch long hook and pile/nylon webbing attachment straps. The straps shall be shipped in the carrying case.
Separate stepdown transformer. The unit shall include a separate stepdown transformer as a separate component not required to be in the carrying case, but shipped in the same shipping package. The transformer shall accept 220-250 volts, 50/60 Hz and provide one half output voltage for the suction apparatus. The transformer shall include a hospital grade, 110 volt receptacle to accept the suction apparatus� specified male line cord. The transformer shall include 3 posts in a compartment with a wire strain-relief attachment for SJT-16/3 wire. The three posts shall be wired to accept any worldwide wiring scheme required to connect a cord for 220-250 volts, single phase AC service. The separate stepdown transformer shall be rated for continuous standby use. The transformer shall comply with UL-1012, including testing.
Power sources. The self-contained apparatus shall be operational from external 12-14 volts DC vehicle generating system rated at 120 amperes at 14 volts, external 120 amp-hour heavy-duty 12 volt battery, 21 volt internal battery, and 110-125 volt, 50/60/400 Hz power sources.
In addition, the separate transformer shall be employed as a power source for the apparatus by reducing 220-250 volts, 50/60 Hz to 110-125 volts.
Operating temperature range. The operating temperature range shall include -4�F to +115�F. The apparatus shall meet the requirements in accordance with ASTM F960 using the internal battery, the 120 amp hour external battery, and 110 volts AC 50, 60, and 400 Hz.�
For Parts Inquires call Parts Hangar, Inc (727) 493-0744
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business