24 January 1992
COMMERCIAL ITEM DESCRIPTION
AUDIOMETER
The General Services Administration has authorized the use of this Commercial Item Description.
This Commercial Item Description covers a microprocessor controlled screening audiometer with integral electrostatic printer.
Salient Characteristics:
General:
1. The audiometer shall comply with all applicable requirements of the following publications, of the issue in effect on the date of solicitation:
a. ANSI S3.6 – Specifications for Audiometers
b. UL 544 – Electric Medical and Dental Equipment
2. Electrical requirements:
a. The audiometer shall be capable of operating over the full tolerance range of one or more of the following line voltages and frequency combinations as specified by the customer requisition:
115VAC plus or minus 10 percent, 50 plus or minus 2Hz.
115VAC plus or minus 10 percent, 60 plus or minus 2Hz.
230VAC plus or minus 10 percent, 50 plus or minus 2Hz.
230VAC plus or minus 10 percent, 60 plus or minus 2Hz.
The dual voltage and hertz feature is desirable but not mandatory.
b. Change over devices, switches, circuits, or transformers if use=d for the purpose of voltage or frequency conversion, shall be clearly marked and identified in the