COMMERCIAL ITEM DESCRIPTION A-A-51036
Unit. Each (EA). One adapter, as specified, constitutes one unit of issue.
Quality Assurance Provisions.
Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contact or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements.
Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the Government’s request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate.
Inspection. Inspection, as used herein, is defined as both examination (such as visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item.
Tests. Tests shall be conducted to determine compliance with specification requirements. Where feasible, the same sample shall be used for the determination of two or more test characteristics.
Sampling for tests. Sampling for tests shall be conducted in accordance with MIL-STD-105 and as specified herein. The unit of product for test purposes shall be one adapter.
Testing requirements. adapters shall be subjected to the tests specified herein.
Corrosion-resistance. The corrosion-resistance of the adapter shall be determined by the Boil Test contained in GG-I-526.
Dimensional testing. The dimensional testing of the adapter shall be in accordance with Figure 1 and ANSI/HIMA MD70.1.
Luer-Lock performance. The adapter Luer-Lock shall be tested in accordance with Section 8 of ANSI/HIMA MD70.1.
For Parts Inquires call Parts Hangar, Inc (727) 493-0744
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business