A-A-51712A
Sampling for tests. Sampling for tests shall be conducted in accordance with MIL-STD-105 and as specified below. The unit of product for test purposes shall be one tong assembly.
Characteristics |
Requirements |
Test procedures |
Inspection level |
AQL |
Composition Analysis |
ASTM F899 |
Requirements |
|
0.0 |
Hardness Test |
ERE 75-100 |
ASTM E 18 E 92 E 140 |
S-2 |
1.5 |
Dimensions |
Figure 1 |
--- |
S-1 |
2.5 |
Copper Sulfate test |
ASTM 1089 |
ASTM 1089 |
S-2 |
1.0 |
Boil test |
ASTM 1089 |
ASTM 1089 |
S-2 |
1.0 |
Test.
Corrosion resistance. The tongs shall be tested as specified in ASTM Designation 1089.
Hardness. Hardness shall be tested in accordance with ASTM E18 or E92 whichever is applicable. Conversion of hardness values shall be in accordance with ASTM E140.
Contractor certification. The contractor shall certify that the product offered meets the salient characteristics of this description and conforms to the producers’ own drawings, specifications, standards, and quality assurance practices. The Government reserves the right to require proof of such conformance prior to first delivery and thereafter as may be otherwise provided for under the provisions of the contract.
Tests. Tests shall be conducted to determine compliance with specification requirements. Where feasible, the same sample shall be used for the determination of two or more test characteristics.