A-A-54038
The apparatus shall include controls to select power on/off and low/high speed, indicating lamps on and battery charger on; circuit breaker/fuse to protect motor from overload; adjustable vacuum valve to regulate maximum vacuum; and gauge to indicate actual vacuum level.
The apparatus shall not include an automatic shut-off. The design of the motor-pump shall be such that overflow of patient debris shall not clog or jam it during use and it shall be readily cleanable.
The apparatus shall operate from its internal batteries (power pack), an external 12 VDC amp-hour heavy-duty vehicle battery, and any 5 amp DC service with voltage within range of 12 to 15 volts.
The dual power source battery changing circuit shall be capable of accepting input power from a vehicle generating system rated 120 amperes for 12 volts system, vehicle 12 volt battery and an AC system of 115 volt, 50/60/400 Hz.
The collection container and tubing shall be sterilizable by ethylene oxide gas or disinfected by cold sterilizing solution. Shall be easily removable without tools for emptying and cleaning.
The c’omplete suction apparatus shall be housed in a rigid or semi-rigid, flame retardant, mildew resistant, plastic case with handle(s), hinge and closure latches. The top of the case shall be permanently marked as follows: EMI/RFI Filtered for Aeromedical Use.
It shall provide user/maintenance manuals (2) which will meet as a minimum ASTM F 960 requirements.
The complete apparatus in its carrying case less the above mentioned manuals, shall not weight more than 20 pounds. The dimensions of the carrying case shall not exceed 20 inches by 15 inches by 12 inches high.
The suction apparatus shall meet the performance requirements listed in this specification.