A-A-54296
14 November 1990
COMMERCIAL ITEM DESCRIPTION
FIXATION DEVICE, EXTERNAL, ORTHOPEDIC
The General Services Administration has authorized the use of this commercial item description.
This commercial item description covers a disposable sterile external fixation system for the immediate stabilization and transportation of patients with open, compound bone fractures.
Salient characteristics:
General. The system shall consist of the following items:
QTY |
ITEM |
1 |
Fixation Rod, Radiolucent, Composite Material, 350 mm Length, with Protective End Caps |
4 or 2 |
4 Adjustable, Single Pin-To-Rod Clamps or 2 Pivoting Double Pin-To-Rod Clamps |
4 |
5.0 mm Threaded Pin, 165 mm Length, Self Tapping and Self Drilling |
1 |
Combination Pin Driving/Clamp Adjusting Wrench Handle |
Rod. Rod shall be fabricated of a radiolucent composite material. The composite shall be graphite/epoxy or carbon fiber/epoxy. The rod shall have a round cross-section and a length of 350 (+/-) mm. When loaded in 3-point bending, the rod shall have a flexural rigidity of 26 Nm2 minimum. The diameter of the rod shall be either 8 (+/- .5) mm or 11 (+/- .5) mm and shall be compatible with the adjustable clamps of the system. The surface texture of the rod shall enable the clamps to securely lock to it. The rod shall have protective end caps and shall be capable of being cut by a gigli saw blade evenly without frayed sharp edges.
DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited.