The fetal heart monitor shall be supplied with the following transducers.
a. Ultrasonic transducer
b. Tocodynomometer transducer
c. Intrauterine pressure transducer
Electrical characteristics.
a. The system shall operate over the full tolerance range of one or more of the following line voltages and frequency combinations
* 115 VAC plus or minus 10 percent, 50 plus or minus 2 Hz
* 115 VAC plus or minus 10 percent, 60 plus or minus 2 Hz
* 230 VAC plus or minus 10 percent, 50 plus or minus 2 Hz
* 230 VAC plus or minus 10 percent, 60 plus or minus 2 Hz
Dual voltage and frequency operation is desirable but is not mandatory.
b. Change over devices, switches, circuits, or transformers if used for the purpose of voltage or frequency conversion, shall be clearly marked and identified in the appropriate documentation which accompanies the unit. Mechanical or electrical adjustment to effect the conversion shall not require the use of specialized tools or skills.
Operational and service data manuals.
a. Information content:
The contractor shall furnish manuals, handbooks and/or brochures containing complete operation, installation and service/maintenance instructions (including pictures of illustrations, as necessary) with complete schematics and wiring diagrams. All manuals, handbooks, and/or brochures shall be written in the English language and all schematics and wiring diagrams shall use American electrical and electronics symbols. The manuals shall include electrical data connection diagrams for all applicable utilities. The instructions shall also contain a complete list of all replaceable parts showing part number, name, and quantity