4. The battery power source shall have the following characteristics:
a. Battery type: Rechargeable
b. Battery capacity: 4 hours of continuous cardioscope monitoring or 1 hour of continuous recording (or any linear combination) at 250C with a HR of 60 BPM and a 1.5 cm cardioscope display.
Nominal thirty (30) discharges of maximum energy at 77°F.
c. Battery meter(s). Shall indicate approximate voltage level of batteries.
d. Battery charge indicator(s): Shall illuminate when batteries are charging.
e. Battery charge time: 16 hours maximum to full charge for each depleted battery.
5. The battery power source shall be installed in the equipment, and in a condition which optimizes shelf-life and minimizes potential damage for the particular battery type.
6. In addition to the requirements specified the contractor shall include specific technical data on battery care and maintenance.
Accessories. Each ECG monitor/recorder and defibrillator shall be supplied with all equipment and accessories required so that each unit comprises a complete operational system. It shall include as a minimum, but not be limited to, the following:
1. One each - Electrode accessory kit to include the following items:
* Four each - Adult limb electrodes.
* One each - Welsh suction cup, 30 mm.
* One each - Welsh suction cup, 15 mm.
* Four each - Adult limb strap, 40 cm.
* One each - Patient cable, 5 lead, 6-pin connector with pin ends for use with metal electrodes.