A-A-54579
One right angled blunt spatula to elevate vitreous and other tissue bridging, or break delicate vitreoretinal adhesions.
The handle and tip of each of the required surgical instruments shall conform to the requirements in figure 1 for dimensions and configuration.
Each required surgical instrument shall be light weight, corrosion-resistant, autoclavable, and designed to be handheld and non-powered. Each required surgical instrument shall be six (6) inches long and shall consist of:
A four (4) inch diamond knurled handle (see figure 1) fabricated from Beta III titanium alloy.
A two (2) inch tapered tip (see figure 1) fabricated from Beta III titanium alloy or 17-4PH corrosion-resisting steel.
Sterilization Case. Shall be supplied with a protective sterilization case that is designed to hold surgical instruments securely in place without inflicting any damage to the tip ends. The sterilization case shall have nominal overall dimensions of 8 inches long X 3 inches wide X 1 inch deep, and shall include the following:
A housing with a low reflective glare, that is fabricated from 22 gauge type 304 corrosion-resisting steel and includes drain holes for condensate collected during a sterilization process (autoclaving).
A hinged corrosion-resisting metal cover with a compression spring catch.
A pivotable metal block that is secured toward an inner edge of the housing (short dimension)
A bracket that is secured within the housing at a location above the pivotable metal block.
All metal parts shall be suitably protected against corrosion.